Comparison of an Adaptive Optics Vision Analyzer and a Hartmann-Shack Aberrometer for measuring ocular wave aberrations

Year: 2014
Type: dissertation
Abstract: Measuring the ocular wavefront aberration function by means of the Hartmann-Shack technique is a common practice in visual optics. Nowadays several commercial devices provide this functionality; this is the reason why it is important that agreement between measurements of different devices exists. In this study a commercial Hartmann-Shack aberrometer (KR-1W, Topcon Corp., Japan) and an Adaptive Optics Vision Analyzer based on a Hartmann-Shack aberrometer (AOVA, Voptica S.L., Spain) have been compared. In order to carry out this study, the wavefront aberration function expressed as a Zernike Polynomials expansion up to the fifth order of 29 right eyes has been obtained. A statistical analysis has been carried out by comparing the individual Zernike coefficients and the radial Root Mean Square (RMSrad) from the second to the fifth order, the high order RMS (RMSHOA) and the total RMS (RMSTOT). The values of the Sphere (S) and Cylinder (C) in Diopters (D) have also been compared. A paired sample t-test, Pearson correlations as well as the Bland and Altman method have been used to assess the agreement between devices. The results showed in general a good agreement between devices when comparing most of the Zernike coefficients.