Year: 1993
Type: article
Abstract: At LEP, a compton polarimeter is used to measure the degree of transverse polarization of the electrons by measuring the mean shift between the vertical distributions of scattered laser light with left and right circular polarization.In this report, the performance of the instrument is discussed.Due to the degree of electron beam polarization being frequently less than 5%, systematic effects producing a mean shift have to be kept below very small values, say about 1 p, which requires a detailed understanding of the instrument.
Authors: R. Aßmann, Jean-Pierre Koutchouk, Rasmus E. Jacobsen, A. Blondel, J. Wenninger, J. Adin Mann, B. Dehning, G. Ferri, M. Placidi, R. Schmidt
Language: English
Subfield: Nuclear and High Energy Physics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W34320165