The optical constants of plutonium metal between .7 and 4.3 eV measured by spectroscopic ellipsometry using a double-windowed experimental chamber.
Year: 2006
Type: article
Abstract: A double-windowed vacuum-tight experimental chamber was developed, and calibrated on the spectroscopic ellipsometer over the energy range from .7 to 4.5 eV using a silicon wafer with approximately 25 nm oxide thickness to remove the multiple-window effects from measurements. The ellipsometric measurements were done such that incident and exit beam were at 65 degree from surface normal. The plutonium sample (3 mm diameter, .1 mm thick) was electro-polished and mounted into the sample chamber in a glove box having a nitrogen atmosphere with less than 100ppm moisture and oxygen content. The index of refraction n and the extinction coefficient k decrease from 3.7 to 1 and 5.5 to 1.1 respectively as the photon energy increases from .7 to 4.3 eV.