Work

Digital behandling av linjärpulser från en CdTe-detektor : en förstudie

Year: 1997
Type: report
Abstract: The neccesary treatment of the linear pulses from a CdTe detector in order to improve the energy resolution for gamma-ray spectroscopy is normally performed by using analogue technique. In this paper we suggest two methods based on digital treatment of the detector pulses. Significant features of the methods are the improvement of the energy resolution, the fact that virtually no dead time is introduced in the detector system and the simpler handling of such systems. The paper describes the underlying idea of the methods, computer simulations of detector system and actual measurements. Preliminary results show that an improvement of the energy resolution of a factor of 2 to 5, depending on the method, used is achieved.
Language: English
Open Access status: green
OpenAlex ID: W36945481
Cites:
Cited by:
Related to: 10