Work

FDG PET/CT for radiotherapy treatment planning. Comparison of functional volume delineation algorithms

Year: 2011
Type: conference-paper
Study design: Observational Study
Abstract: 1792 Objectives PET/CT imaging could improve delineation of rectal carcinoma gross tumor volume (GTV) and reduce inter-observer variability. Our objective is to compare different functional volume delineation algorithms. Methods We enrolled 33 consecutive patients with Locally Advanced Rectal Carcinoma. The FDG PET/CT and the high dose CT (CTRT) were performed in the radiation treatment position. For each patient, the GTVRT was delineated based on the CTRT and compared to 5 different GTVPET derived from an automatic segmentation approach (FLAB) [3]; a relative threshold (45% of the pixel max) and an absolute threshold (SUV>2.5), using 2 commercially available softwares (Philips EBW4 and Segami OASIS). Results were analyzed using Friedman Test. Results For the all group, the tumor volumes were as follow (mean ± SD): GTVRT 39.4 ± 30.9 ml; FLAB 21.5 ± 18 ml; Oasis 45% 15.8 ± 14.2 ml; Philips 45% 16.4 ± 12.4 ml; Oasis 2.5 40.8 ± 30.7 ml; Philips 2.5 42.4 ± 35.3 ml. Except for Oasis 2.5 and Philips 2.5, all GTVPET volumes were statistically smaller than the GTVRT (p Conclusions We observed large variations in tumor volume delineation according to the segmentation algorithm applied, emphasizing the need for delineation standards
Language: English
Subfield: Radiation
Open Access status: green
OpenAlex ID: W42155295
FWCI:
Cites:
Cited by:
Related to: 10