Work

A system calibration and fast iterative reconstruction method for next-generation SPECT imagers

Year: 2011
Type: conference-paper
Abstract: Recently, high-resolution gamma cameras have been developed with detectors containing >; 105-106elements. SPECT imagers based on these detectors usually also have a large number of voxel bins and therefore face memory storage issues for the system matrix when performing fast tomographic reconstructions using iterative algorithms. To address these issues, we have developed a method that parameterizes the detector response to a point source and generates the system matrix on the fly during MLEM or OSEM on graphics hardware. The calibration method, interpolation of coefficient data, and reconstruction results are presented in the context of a recently commissioned small-animal SPECT imager, called FastSPECT III.
Language: English
Field: Medicine
Open Access status: green
OpenAlex ID: W4229744964
FWCI: 2.598
Cites: 18
Cited by: 5
Related to: 10