A reply

Year: 2007
Type: article
Abstract: Thank you for the opportunity to reply to the letter from Drs Puttick and Malik. The connector in the affected catheter mount was occluded as a result of a moulding fault. As soon as this was reported to us we inspected the remaining stock (over 5000) pieces and found no further occlusions. We conclude that this was an isolated case. Intersurgical aims to detect and prevent blockages in all the respiratory products we manufacture through our testing procedures. Unfortunately, this catheter mount was not detected and, as a result, we have now fully automated the testing procedure, eliminating the possibility of error. Finally, we endorse the practice of carrying out a visual and a flow test of all components of anaesthetic breathing systems, as recommended by the AAGBI.