Year: 2004
Type: letter
Abstract: Syringe production failureWe report a rare incident whereby the quality assurance processes of a major manufacturer of syringes failed.After opening a 20 ml BD Plastipak Syringe (Becton Dickinson & Co; Drogheda, IRELAND) Lot 7503A10 Exp 2008-01 it was immediately noted that a component of the mould had been left in the outlet tip during manufacture.(See Figures 1 and 2.) Since this had separated during production, colleagues hypothesized that other syringes might have been produced with blocked outlets in subsequent batches.This turned out to be the case a month later, when a second syringe appeared but unfortunately it was discarded by the anaesthetist before being photographed.A report was lodged with the Therapeutic Goods Administration under its Australian and New Zealand Medical Device Incident Report Investigation Scheme in September 2003.The company reported back that "no other reports of this nature had occurred.It was due to a break in the bonding material, which holds the tip pin to the mould.Syringes which followed this one had a blocked tip and during the inspection process were detected and disposed of.This syringe however evaded detection".Because there were no other reports of this problem and no patient injury occurred, the TGA determined not to investigate further.