Year: 2021
Type: article
Abstract: Calibration of the gain and digital conversion factor of an EMCCD is necessary for accurate photon counting. We present a new method to quickly calibrate multiple gain settings of an EMCCD camera. Acquiring gain-series calibration data and analyzing the resulting images with the EMCCD noise model more accurately estimates the gain response of the camera. Furthermore, we develop a method to compare the results from different calibration approaches. Gain-series calibration outperforms all other methods in this self-consistency test.
Source: Scientific Reports
Language: English
Topic: CCD and CMOS Imaging Sensors
Subfield: Electrical and Electronic Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $1,990
OpenAlex ID: W3199625172