Year: 2010
Type: article
Abstract: We describe the results of a comparison of reference standards between the National Institute of Standards and Technology (NIST-USA) and National Institute of Metrology (NIM-China). We report optical fiber-based power measurements at nominal wavelengths of 1310 nm and 1550 nm. We compare the laboratories' reference standards by means of a commercial optical power meter. Measurement results showed the largest difference of less than 2.6 parts in 10(3), which is within the combined standard (k = 1) uncertainty for the laboratories' reference standards.
DOI: 10.6028/jres.115.029
Authors: I. Vayshenker, D.J. Livigni, J.H. Lehman, J. Li, L.M. Xiong, Z.X. Zhang
Language: English
Subfield: Instrumentation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: bronze
OpenAlex ID: W4249540905