Work

Retraction statement: Dental hard tissue ablation using mid‐infrared tunable nanosecond pulsed Cr:CdSe laser

Year: 2018
Type: retraction
Abstract: Retraction: The article “Dental hard tissue ablation using mid-infrared tunable nanosecond pulsed Cr:CdSe laser” by Taichen Lin DDS, PhD, Akira Aoki DDS, PhD, Norihito Saito PhD, Masaki Yumoto PhD, Sadahiro Nakajima MS, Keigo Nagasaka PhD, Shizuko Ichinose PhD, Koji Mizutani DDS, PhD, Satoshi Wada PhD, and Yuichi Izumi DDS, PhD published in Lasers in Surgery and Medicine online on March 29, 2016 in the Wiley Online Library (www.wileyonlinelibrary.com), has been retracted by agreement between the authors, the journal Editor-in-Chief, and Wiley Periodicals, Inc. because there was a systemic error in the energy output of each wavelength used in the study. The sensor of the power meter had a systemic problem (some wavelengths near 2.7un were absorbed by the surface material [diffuser] or the sensor, which was not recognized by the production company itself). Consequently, the data of figures three through seven were not correct and the conclusion became completely different. Reference Lin T, Aoki A, Saito N, Yumoto M, Nakajima S, Nagasaka K, Ichinose S, Mizutani K, Wada S, Izumi Y. Dental hard tissue ablation using mid-infrared tunable nanosecond pulsed Cr:CdSe laser. Lasers Surg. Med. 2016;48:965–977. https://doi.org/10.1002/lsm.22508
Language: English
Field: Medicine
Open Access status: bronze
OpenAlex ID: W4256431404
Cites: 1
Cited by:
Related to: 10