T1-weighted Black-Blood Carotid Plaque Imaging Using Variable Flip-Angle 3D Turbo Spin-Echo : Comparison with 2D Turbo Spin Echo and 3D Turbo Field Echo Sequences
Year: 2011
Type: article
Abstract: :Purpose: To compare a T1-weighted, three-dimensional variable-flip-angle (3D-VRFA) turbo spin-echo (TSE) sequence with both a two-dimensional (2D)-TSE double inversion recovery (DIR) sequence and a 3D turbo field-echo (3D-TFE) DIR sequence in magnetic resonance black-blood carotid plaque imaging.Materials and Methods: Fifteen patients underwent black-blood plaque imaging including pulse-triggered axial T1-weighted (T1W) 2D-TSE, pulse-triggered sagittal 3D-TFE, and non-triggered coronal T1W 3D-VRFA sequences.The imaging quality of each sequence was visually graded on a three-point score.The signal ratio of the carotid plaque to the arterial lumen (RPL), and that of the carotid plaque to the adjacent muscle (RPM) was calculated in each sequence.Results: The score for the 3D-TFE was lower than that of 3D-VRFA.Although no significant differences were observed between the scores of the 2D-TSE and 3D-VRFA sequences, the 3D-VRFA allowed visualization of arteries in arbitrary orientations, as well as small plaque compositions such as ulcerations and calcifications not visualized on 2D-TSE.The RPL was highest on 2D-TSE, whereas the RPM was highest on 3D-VRFA.Conclusions: 3D-VRFA is a promising technique for the diagnosis of carotid plaques.