Work

Evaluating the Impact of Applying Personal Glucose Targets in a Closed-Loop System for People With Type 1 Diabetes

Year: 2022
Type: article
Abstract: Background CamAPS FX is a hybrid closed-loop smartphone app used to manage type one diabetes. The closed-loop algorithm has a default target glucose of 5.8 mmol/L (104.5 mg/dL), but users can select personal glucose targets (adjustable between 4.4 mmol/L and 11.0 mmol/L [79 mg/dL and 198 mg/dL, respectively]). Method In this post-hoc analysis, we evaluated the impact of personal glucose targets on glycemic control using data from participants in five randomized controlled trials. Results Personal glucose targets were widely used, with 20.3% of all days in the data set having a target outside the default target bin (5.5-6.0 mmol/L [99-108 mg/dL]). Personal glucose targets >6.5 mmol/L (117 mg/dL) were associated with significantly less time in target range (3.9-10.0 mmol/L [70-180 mg/dL]; 6.5-7.0 mmol/L [117-126 mg/dL]: mean difference = −3.2 percentage points [95% CI: −5.3 to −1.2; P < .001]; 7.0-7.5 mmol/L [126-135 mg/dL]: −10.8 percentage points [95% CI: −14.1 to −7.6; P < .001]). Personal targets >6.5 mmol/L (117 mg/dL) were associated with significantly lower time (<3.9 mmol/L [<70 mg/dL]; 6.5-7.0 mmol/L [117-126 mg/dL]: −1.85 percentage points [95% CI: −2.37 to −1.34; P < .001]; 7.0-7.5 mmol/L [126-135 mg/dL]: −2.68 percentage points [95% CI: −3.49 to −1.86; P < .001]). Conclusions Discrete study populations showed differences in glucose control when applying similar personal targets.
Language: English
Field: Medicine
Open Access status: hybrid
Estimated APC paid: $3,500
OpenAlex ID: W4312065417
FWCI: 0.741
Cites: 8
Cited by: 4
Related to: 10