Comparison of continuous air monitor utilization: A case study
Year: 1997
Type: article
Abstract: The Chemical Metallurgy Research (CMR) building has been upgrading to different continuous air monitors (CAMs) over the past several years. During the transition, both the newer and older CAMs were positioned in the rooms for field testing and comparison. On December 19, 1996, an accidental release of plutonium aerosol occurred into a laboratory in the CMR building. The event occurred while the room was unoccupied, and no personnel were exposed from this incident. There were two fixed air samplers (FASs) and three CAMs operating in the room at the time the release occurred, including two of the recently installed Canberra Alpha Sentry CAMs and one older Eberline CAM. The apparent cause of the release was a procedure carried out in the basement involving the replacement of the HEPA filter in the ventilation exhaust of a slot-box in the laboratory. For a short period, the ventilation was disconnected from the slot-box in this room, but not from the chemical hood exhaust on the opposite side of the laboratory. Therefore, a condition was created where backflow could occur out of the slot-box and into the room. Eventually all three CAMs in the room alarmed, and the situation was successfully monitored and brought under control by health physics personnel. Data on CAM performance were logged, and Pu activity collected on CAM and FAS filters were measured. A comparison of the new and old continuous air monitoring programs was performed and many interesting lessons on CAM performance and CAM utilization were learned. Overall, this comparison showed the advantages of remote monitoring, timely spectral information, and concentration measurements resolved in time and space.
Authors: John Rodgers, Jeffrey Whicker, J.T. Voss
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
Funder: U.S. Department of Energy
OpenAlex ID: W43349929