Seeing Is Believing: Photon Counting Computed Tomography Clearly Images Directional Deep Brain Stimulation Lead Segments and Markers After Implantation
Year: 2023
Type: article
Study design: Observational Study
Abstract: BACKGROUND AND OBJECTIVES: Directional deep brain stimulation (DBS) electrodes are increasingly used, but conventional computed tomography (CT) is unable to directly image segmented contacts owing to physics-based resolution constraints. Postoperative electrode segment orientation assessment is necessary because of the possibility of significant deviation during or immediately after insertion. Photon-counting detector (PCD) CT is a relatively novel technology that enables high resolution imaging while addressing several limitations intrinsic to CT. We show how PCD CT can enable clear in vivo imaging of DBS electrodes, including segmented contacts and markers for all major lead manufacturers. MATERIALS AND METHODS: We describe postoperative imaging and reconstruction protocols we have developed to enable optimal lead visualization. PCD CT images were obtained of directional leads from the three major manufacturers and fused with preoperative 3T magnetic resonance imaging (MRI). Radiation dosimetry also was evaluated and compared with conventional imaging controls. Orientation estimates from directly imaged leads were compared with validated software-based reconstructions (derived from standard CT imaging artifact analysis) to quantify congruence in alignment and directional orientation. RESULTS: High-fidelity images were obtained for 15 patients, clearly indicating the segmented contacts and directional markers both on CT alone and when fused to MRI. Our routine imaging protocol is described. Ionizing radiation doses were significantly lower than with conventional CT. For most leads, the directly imaged lead orientations and depths corresponded closely to those predicted by CT artifact-based reconstructions. However, unlike direct imaging, the software reconstructions were susceptible to 180° error in orientation assessment. CONCLUSIONS: High-resolution photon-counting CT can very clearly image segmented DBS electrode contacts and directional markers and unambiguously determine lead orientation, with lower radiation than in conventional imaging. This obviates the need for further imaging and may facilitate anatomically tailored directional programming.
Authors: James Manfield, Sheena Thomas, Marko Bogdanovic, Nagaraja Sarangmat, Charalambos Antoniades, Alexander L. Green, James J. FitzGerald
Institutions: John Radcliffe Hospital, University of Oxford
Language: English
Subfield: Neurology
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $3,500
Awards: Developing novel biomarkers for cardiovascular risk stratification using computed tomography imaging of perivascular adipose tissue, 11βHSD1 inhibition: a distinctive acute-MI intervention for promotion of peri-infarct vascularisation, reduction of infarct expansion and prevention of heart failure, G8789181138
OpenAlex ID: W4388297382