Le temps et le mouvement en Physique Médicale

Year: 2018
Type: dissertation
Abstract: Purpose: To assess the validity of gated radiotherapy of lung by using a cross-check methodology based on four-dimensional (4D)-computed tomography (CT) exams.Variations of volume of a breathing phantom was used as an indicator.Methods and Materials: A balloon was periodically inflated and deflated by a medical ventilator.The volume variation (⌬V) of the balloon was measured simultaneously by a spirometer, taken as reference, and by contouring 4D-CT series (10 phases) acquired by the real-time position management system (RPM).Similar cross-comparison was performed for 2 lung patients, 1 with free breathing (FB), the other with deep-inspiration breath-hold (DIBH) technique.Results: During FB, ⌬V measured by the spirometer and from 4D-CT were in good agreement: the mean differences for all phases were 8.1 mL for the balloon and 10.5 mL for a patient-test.End-inspiration lung volume has been shown to be slightly underestimated by the 4D-CT.The discrepancy for ⌬V between DIBH and end-expiration, measured from CT and from spirometer, respectively, was less than 3%.Conclusions: Provided that each slice series is correctly associated with the proper breathing phase, 4D-CT allows an accurate assessment of lung volume during the whole breathing cycle (⌬V error <3% compared with the spirometer signal).Taking the lung volume variation into account is a central issue in the evaluation and control of the toxicity for lung radiation treatments.