Transcatheter Extraction of a Migrated Left Atrial Appendage Occluder Device
Year: 2024
Type: article
Study design: Case Report
Abstract: A 69-year-old man presented at 10 weeks postimplantation with a 31-mm Watchman FLX migrating into the left atrium. Due to incomplete left atrial appendage seal and embolization risk, transcatheter device extraction was performed without complications. Herein we describe the technique and procedural steps, using cardiac computed tomography and benchtop models to guide practice.
Source: JACC Case Reports
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $850
OpenAlex ID: W4401385984