On High Performance Control of Concentric Tube Continuum Robots Through Parsimonious Calibration
Year: 2024
Type: article
Abstract: Continuum robots deform continuously, compared to conventional robots composed of rigid links and joints, and require dedicated calibration methods. Indeed, calibration is an essential step to obtain high performance control, as it directly influences robot accuracy. In this letter, we investigate how model parameters influence both model accuracy and model-based closed-loop control accuracy of Concentric Tube Continuum Robots (CTCR). A fast, robust, and real-time implementation of the Cosserat rod model is first introduced. Then, a model-based Jacobian control scheme is presented. A parsimonious calibration procedure focused on control accuracy is finally proposed to achieve submillimetric tracking errors along a 3D trajectory at velocity reaching 5 mm/s in complex scenarios including actuation constraints, obstacle avoidance, and external forces. Results are demonstrated both in simulation and on an experimental setup of a 3-tube CTCR.
Authors: Quentin Boyer, Sandrine Voros, Pierre Roux, François Marionnet, Kanty Rabenorosoa, M. Taha Chikhaoui
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green
Funder: Agence Nationale de la Recherche
Awards: Gestes Médico-Chirurgicaux Assistés par Ordinateur, MicroRobot-assisted Cholesteatoma Surgery, G2659019386, G2776477858, Infrastructure technologique pour la recherche d’excellence en robotique, Ingénierie et Innovation par les sciences physiques, les savoir-faire technologiques et l'interdisciplinarité, G6803667113, COSserat theory for SlEnder RObOTS
OpenAlex ID: W4402299963