Development of a three‐dimensional scintillation detector for pencil beam verification in proton therapy patient‐specific quality assurance
Year: 2024
Type: article
Abstract: BACKGROUND: Pencil Beam Scanning proton therapy has many advantages from a therapeutic point of view, but raises technical constraints in terms of treatment verification. The treatment relies on a large number of planned pencil beams (PB) (up to thousands), whose delivery is divided in several low-intensity pulses delivered a high frequency (1 kHz in this study). PURPOSE: The purpose of this study was to develop a three-dimensional quality assurance system allowing to verify all the PBs' characteristics (position, energy, intensity in terms of delivered monitor unit-MU) of patient treatment plans on a pulse-by-pulse or a PB-by-PB basis. METHODS: scintillator cube and a fast camera, synchronized with beam delivery, recording two views (direct and using a mirror) of the scintillation distribution generated by the pulses. A specific calibration and analysis process allowed to extract the characteristics of all the pulses delivered during the treatment, and consequently of all the PBs. The system uncertainties, defined here as average value + standard deviation, were characterized with a customized irradiation plan at different PB intensities (0.02, 0.1, and 1 MU) and with two patient's treatment plans of three beams each. The system's ability to detect potential treatment delivery problems, such as positioning errors of the treatment table in this work (1° rotations and a 2 mm translation), was assessed by calculating the confidence intervals (CI) for the different characteristics and evaluating the proportion of PBs within these intervals. RESULTS: MU) allowing uncertainties smaller than 580 µm for the position, 180 keV for the energy and 3% for the intensity on patients treatment plans. The position and energy uncertainties were found to be little dependent from the pulse intensities whereas the intensity uncertainty depends on the pulses number and intensity distribution. Finally, treatment plans evaluations showed that 98% of the PBs were within the CIs with a nominal positioning against 83% or less with the table positioning errors, thus proving the ability of SCICOPRO to detect this kind of errors. CONCLUSION: MU. SCICOPRO was thus able to measure all the characteristics of the spots of a treatment (position, energy, intensity) in a single measurement, making it possible to verify their compliance with the treatment plan. SCICOPRO thus proved to be a fast and accurate tool that would be useful for patient-specific quality assurance (PSQA) on a pulse-by-pulse or PB-by-PB verification basis.
Source: Medical Physics
DOI: 10.1002/mp.17388
Institutions: Centre National de la Recherche Scientifique, Commissariat à l'Énergie Atomique et aux Énergies Alternatives, Institut National de Physique Nucléaire et de Physique des Particules, Grand Accélérateur National d'Ions Lourds, Normandie Université, Université de Caen Normandie, École Nationale Supérieure d'Ingénieurs de Caen, Laboratoire de Physique Corpusculaire de Caen, Centre François Baclesse
Language: English
Subfield: Pulmonary and Respiratory Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $3,200
Funder: Région Normandie
OpenAlex ID: W4402402153