Impact of commissural alignment on the hemodynamic performance of supra‐annular self‐expandable transcatheter aortic valves

Year: 2024
Type: article
Study design: Observational Study
Abstract: ABSTRACT Background Impact of commissural alignment (CA) strategies with self-expandable supra-annular transcatheter aortic valves (TAVR) has not been investigated yet. Methods Multicentric ambispective study including patients who underwent self-expandable TAVR in 7 centers with the Evolut Pro/Pro+ (EP)(Medtronic, USA) and Acurate neo2 (AN2)(Boston Scientific, USA). Comparison of those with CA attempt vs. those with no CA attempt was performed. The degree of commissural misalignment (CMA) was assessed by computed tomography/angiography and 1-year transvalvular gradients/regurgitation evaluated by echocardiography. A matched comparison according to annular dimensions/eccentricity, prosthesis size/type, and baseline left ventricular function and gradients was performed. Results A total of 557 patients, mean age 80.7±6.6 years, 61.4% men, and STS score of 4.3±3.1% were analyzed. A CA technique was attempted in 215 patients (38.6%), including 113 patients with AN2 and 102 patients with EP. None/mild CMA was found in 158 (73.5% vs. 43.6% if no CA attempted, p<0.001) with no differences between devices (AN2:75.2%; EP:71.6%, p=0.545). Patients with moderate/severe CMA had a greater aortic peak gradient (22.3±8.7 vs. 19.7±8.5, p=0.001), significantly greater progression of both peak (p=0.002) and mean gradients (p=0.001) after matching, and higher rate of central aortic regurgitation (1.2% vs. 0.4%, p=0.005) at 1-year, but not a greater proportion of patients with mean gradient ≥10mmHg. Conclusions The use of CA strategies significantly reduced the rate of CMA for the self-expandable TAVR devices ACN2 and EP which was associated to lower transvalvular gradients and intra-prosthetic regurgitation progression at 1-year although no criteria of structural deterioration were met at this follow up. Clinicaltrials.org: NCT05097183 CONDENSED ABSTRACT Commissural alignment (CA) strategies with Evolut Pro and Acurate neo2 TAVR devices were similarly successful in ¾ of the patients, representing a significant increase compared to traditional implantation. Patients with moderate/severe commissural misalignment (CMA) had a significantly greater progression of both peak and mean gradients and higher rate of new central aortic regurgitation at 1-year, but not a greater proportion of patients with mean gradient above 10mmHg. Hence, even at only 1-year of follow up and despite the lack of formal structural deterioration criteria, some hints suggest a more rapid deterioration if moderate/severe CMA is present after supra-annular TAVR.
Language: English
Field: Medicine
Open Access status: green
OpenAlex ID: W4402419396
FWCI: 1.069
Cites: 19
Cited by: 4
Related to: 10