Leg occlusion of the Ovation Alto stent graft main body caused by twisting of the stent graft: A case report
Year: 2024
Type: article
Study design: Case Report
Abstract: The Ovation Alto stent graft is a unique endovascular abdominal aortic aneurysm repair device. A low-viscosity, radiopaque liquid polymer fills the rings located within the legs of the device. We reported a rare case of a patient diagnosed with an abdominal aortic aneurysm who was surgically treated using endovascular abdominal aortic aneurysm repair with an Ovation Alto stent graft. Polymer underfilling due to contralateral leg twisting of the stent graft's main body resulted in limb occlusion. We highlighted the importance of avoiding the rotation of the Ovation Alto stent graft, thus preventing the potential twisting of this device within the sheath after the insertion of the stent graft's main body.
Authors: Masaaki Naganuma, Yukihiro Hayatsu, Ryoichi Tsuruhara, Hayate Nomura, Naoya Terao, Kazuhiro Yamaya, Masaki Hata
Institution: Sendai Kousei Hospital
Language: English
Subfield: Pulmonary and Respiratory Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $1,250