Pharmacokinetics of Subcutaneous Itepekimab Injection With an Autoinjector Device and Prefilled Syringe in Healthy Participants
Year: 2024
Type: article
Study designs: Randomized Controlled Trial, Clinical Trial
Abstract: Itepekimab, a monoclonal antibody against interleukin-33, has demonstrated clinical utility in previous studies in patients with asthma and chronic obstructive pulmonary disease. An autoinjector (AI) has been developed for administering itepekimab to facilitate further development. This study compared pharmacokinetics of single 300-mg itepekimab subcutaneous administration via an AI versus a prefilled syringe (PFS). Of 90 healthy volunteers enrolled in this Phase 1, parallel-design, randomized study and stratified by body weight (50 to <70 kg, ≥70 to <80 kg, ≥80 to 100 kg) and injection site (abdomen, thigh, or arm), 84 completed the study. Systemic exposure of itepekimab was similar for both groups. Point estimates for geometric mean ratios of pharmacokinetic parameters for AI versus PFS groups were 1.01 for maximum serum concentration, 1.06 for area under the serum concentration-time curve to the last quantifiable concentration, and 1.04 for area under the serum concentration-time curve extrapolated to infinity. The exposure was similar for both devices in each body weight and injection site subgroup. Overall, systemic exposure of 300-mg single-dose itepekimab in healthy participants was comparable when administered subcutaneously via an AI device and PFS, with an acceptable safety profile in both device groups.
DOI: 10.1002/cpdd.1466
Authors: Christine Xu, Kong Xin, Matthew P. Kosloski, Allison Butler, Helene Goulaouic, Michael C. Nivens, Vanaja Kanamaluru
Language: English
Subfield: Immunology
Field: Immunology and Microbiology
Domain: Life Sciences
Open Access status: hybrid
Estimated APC paid: $3,620
Funders: Sanofi, Regeneron Pharmaceuticals
OpenAlex ID: W4402725697