Work

Room temperature dual-mode internal quantum deficiency measurement with propagated uncertainty to 0.03% (k = 2)

Year: 2024
Type: article
Abstract: We present improvements in dual-mode calibration of predictable quantum efficient detectors and demonstrate the importance of calculating absolute uncertainties instead of relative uncertainties. We have implemented a new uncertainty component for the thermal fluctuations in the temperature signal which results in a propagated Type A uncertainty, matching the observed standard deviation. A new thermal drift correction method exploiting a monitor thermistor on the heat sink is relaxing the need for thermal stabilisation of the experimental set-up. With beam position uncertainty ±0.25 mm and background electrical power varying from 10 μW to 900 μW, the measured internal quantum deficiency (IQD) is in average 0.00% ± 0.03% (k = 2). The IQD exhibits clear systematic effects of beam position and background power, showing the need for improved design of dual-mode modules to further improve the uncertainty.
Source: Metrologia
Language: English
Open Access status: hybrid
Estimated APC paid: $3,165
Award: G377608323
OpenAlex ID: W4402803267
FWCI: 0.178
Cites:
Cited by: 1
Related to: 10