Verification of analytical performance of intact parathyroid hormone assay on the Abbott Alinity ci®: Experience of the central laboratory of Mohammed VI University Hospital of Oujda

Year: 2024
Type: article
Abstract: The primary goal of our study was to verify the analytical performance of intact parathyroid hormone measurements conducted on the Abbott CI analyzer using the immuno-chemiluminescence method. This verification took place in the biochemistry laboratory at Mohammed VI University Hospital in Oujda. Our methodology was aligned with the recommendations outlined in the French accreditation committee (COFRAC) technical guide (GTA) 04, emphasizing the assessment of reproducibility and repeatability. Overall, the results of this evaluation were satisfactory and met the standards set by both the supplier and the French Society of Clinical Biology. This study illustrates that the biochemistry laboratory at Mohammed VI University Hospital of Oujda is equipped to provide accurate and precise results, which are crucial for effective clinical diagnosis and decision-making.