Year: 2024
Type: article
Study design: Observational Study
Abstract: OBJECTIVE: To evaluate the accuracy of diagnosing dry eye disease (DED) by using the Visionix VX120+, to establish reference values for tear meniscus height (TMH) and non-invasive break-up time (NIBUT), and to compare the NIBUT measurements with the fluorescein tear break-up time (FBUT), Methods: fifty-eight subjects (34 dry eye and 24 control) were enrolled. The TMH, first NIBUT, and NIBUT50% were evaluated with the Visionix VX120+, and the FBUT was measured with a slit-lamp. The Receiver Operating Characteristic (ROC) curve was used to evaluate the diagnostic performance, and the Bland-Altman method was performed to analyze the agreement. RESULTS: The areas under the curve were 0.62, 0.60, and 0.70 for the TMH, first NIBUT, and NIBUT50%, respectively. The optimal cut-off values (sensitivity, specificity) were 0.29 (0.62, 0.67), 5.05 (0.85, 0.46), and 7.35 (0.65, 0.79) for the TMH, first NIBUT, and NIBUT50%, respectively. The mean differences (lower, upper limits of agreement) were -1.10 (-8.78, 6.58) and 1.55 (-5.68, 8.78) for the first NIBUT vs. FBUT and the NIBUT50% vs. FBUT, respectively. CONCLUSIONS: In conclusion, the NIBUT50% can be a useful tool for dry eye screening, with acceptable values of sensitivity and specificity. First, the NIBUT and NIBUT50% should not be used interchangeably with the FBUT.
Source: Diagnostics
Institutions: Universidad de Valladolid, University of Alicante
Language: English
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $2,955
Award: G8407169708
OpenAlex ID: W4403401105