Year: 2024
Type: article
Abstract: The National Ignition Facility produced the first nuclear fusion experiment demonstrating net positive energy gain on December 5, 2022. The x-ray streak camera that measures the bang time and burn-width from this landmark experiment had an electronic failure and did not record data. The CCD sensor was replaced with a radiation hardened CMOS sensor that has since demonstrated successful operation on repeat ignition shots. Concurrently, an instrument artifact was identified that occurs when the signal consists primarily of energetic x rays >15 keV (common on burning plasma experiments). This artifact, which appears as a background pedestal, arises from the x-ray back-fluorescence generated by the solid metal accelerating mesh behind the photocathode in the streak tube. We have mitigated this background signal by limiting the sensitive area of the photocathode. Herein, the details of the modifications and the results are presented.
Source: Review of Scientific Instruments
DOI: 10.1063/5.0219554
Authors: S. F. Khan, P. R. Nyholm, K. J. Decker, A. G. MacPhee, T. J. Hilsabeck, D. Boyle, N. Palmer, T. A. Miller, A. C. Carpenter
Language: English
Subfield: Nuclear and High Energy Physics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
Awards: G1356433355, G14554683, G2145270357, G991281326
OpenAlex ID: W4403406550