CT-Derived Patient-Specific Computer Simulation of the Novel Self-Expanding Evolut FX Implantation: A Case Series
Year: 2024
Type: article
Study design: Case Report
Abstract: Background/Objectives: Paravalvular leak and permanent pacemaker implantation remain relevant issues after transcatheter aortic valve implantation (TAVI). Novel device development as well as adequate preprocedural device selection can contribute to optimal outcomes. Methods: Computed tomography-based patient-specific computer anatomical analysis and simulation were used in addition to standard preprocedural preparation in three of the first Evolut FX cases in our center. Procedural and follow-up echocardiographic outcomes are presented. Results: Computed tomography-based computer simulation of Evolut FX resulted in implantation of a different size in one of three cases. In three cases of severe aortic valve stenosis, procedural as well as follow-up outcomes were satisfactory and in line with the simulated results of the chosen strategy. Conclusions: Preprocedural patient-specific computer simulation predicts and guides decision-making in TAVI with the Evolut FX platform. The combination of advanced preprocedural technology and novel technologies continues to contribute to the optimization of TAVI outcomes.
Source: Journal of Clinical Medicine
DOI: 10.3390/jcm13206212
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $2,955
Award: G8774823627