In vivo assessment of a modification of a domiciliary ventilator which reduces oxygen consumption in mechanically ventilated patients
Year: 2024
Type: article
Abstract: There are instances where hospitals may experience a relative shortage of oxygen, for example, future pandemics, natural disasters or wartime. We developed a modification to a domiciliary ventilator that captures oxygen normally vented during expiration into the atmosphere, delivering it to the patient. The modification significantly increases PaO 2 in mechanically ventilated patients at their baseline FiO 2 and baseline FiO 2 +1, +2 and +3 L/min compared to no modification. The modification reduces walled oxygen consumption by median (IQR) −1.0 (−1.25 to −1.00) L/min whilst maintaining PaO 2 . This modification may be a valuable in circumstances where there is a relative shortage of oxygen.
Authors: Timothy O Jenkins, Thomas M Sutton, Peter Griffen, Yoseph Mebrate, Michael I Polkey
Language: English
Subfield: Pulmonary and Respiratory Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W4404083376