Work

Vacuum-ultraviolet light source for enhanced production of metastable krypton and xenon beams

Year: 2024
Type: article
Abstract: We demonstrate excitation of metastable krypton and xenon beams using a vacuum ultraviolet lamp and directly compare the performance of this method to metastable excitation based on a radio-frequency-driven plasma discharge. In our apparatus, lamp-based metastable excitation outperforms the plasma discharge across a wide range of beam flux values relevant for atom trap trace analysis (ATTA). Moreover, we do not observe significant degradation in lamp performance after over 160 h of operation. We find that lamp-based excitation is particularly advantageous at the smallest and largest beam fluxes tested, demonstrating the utility of this approach both for improving krypton ATTA and for enabling the detection of radioactive xenon isotopes using ATTA. Finally, we demonstrate an additional enhancement to lamp-based metastable excitation efficiency and stability by applying an external magnetic field.
Language: English
Open Access status: green
OpenAlex ID: W4404319824
FWCI: 0.159
Cites: 42
Cited by: 1
Related to: 10