Work

Luminance calibration for DICOM GSDF

Year: 2024
Type: article
Abstract: Accurate image reproduction is important in display technologies. For instance, in medical imaging, correct image reproduction is essential to make accurate diagnoses and treatments. For this reason, the medical community has produced the DICOM (digital imaging and communications in medicine) standard. In this paper, we propose a luminance (gray) calibration method in accordance with the DICOM standard. Theoretically, the proposed method would provide the optimal solution for the DICOM GSDF (grayscale standard display function), though it is not possible to implement the optimal solution due to the time variances of displays and light measuring devices. Experimental results show that the proposed calibration method provides excellent performance for gray calibration.
Authors: Chulhee Lee, Donghun Seok, Munju Kim, Junesung Ko, Daewon Kim, Donguk Kim, Jin Jeon, Semi Lee
Language: English
Open Access status: gold
Estimated APC paid: $2,270
OpenAlex ID: W4404398938
FWCI:
Cites: 10
Cited by:
Related to: 10