Luminance calibration for DICOM GSDF
Year: 2024
Type: article
Abstract: Accurate image reproduction is important in display technologies. For instance, in medical imaging, correct image reproduction is essential to make accurate diagnoses and treatments. For this reason, the medical community has produced the DICOM (digital imaging and communications in medicine) standard. In this paper, we propose a luminance (gray) calibration method in accordance with the DICOM standard. Theoretically, the proposed method would provide the optimal solution for the DICOM GSDF (grayscale standard display function), though it is not possible to implement the optimal solution due to the time variances of displays and light measuring devices. Experimental results show that the proposed calibration method provides excellent performance for gray calibration.
Source: Optics Express
DOI: 10.1364/oe.540462
Authors: Chulhee Lee, Donghun Seok, Munju Kim, Junesung Ko, Daewon Kim, Donguk Kim, Jin Jeon, Semi Lee
Language: English
Subfield: Atomic and Molecular Physics, and Optics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $2,270
Funders: Yonsei University, LG Electronics
OpenAlex ID: W4404398938