A Case of Very Early Lead Fracture in an Implantable Cardioverter‐Defibrillator: Management and Follow‐Up
Year: 2024
Type: article
Study design: Case Report
Abstract: INTRODUCTION: Implantable cardioverter defibrillators (ICDs) lead fractures are rare but serious complications, often leading to inappropriate shocks. The early occurrence of such fractures post-implantation is exceptionally uncommon. METHODS AND RESULTS: We present a case of a 53-year-old male with hypertrophic cardiomyopathy who experienced an inappropriate shock due to a lead fracture just 4 days after ICD implantation. The fractured lead was successfully extracted and replaced using a gooseneck snare passed through an oversized TightRail with no immediate complications. CONCLUSION: Early lead fractures can occur shortly after ICD implantation, leading to inappropriate shocks. Successful management, including extraction and reimplantation, is possible even in these early cases. Clinicians should remain vigilant, and remote monitoring can play a crucial role in the timely detection of such complications.
DOI: 10.1111/pace.15103
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W4404494647