Work

Accurate patient alignment without unnecessary imaging using patient-specific 3D CT images synthesized from 2D kV images

Year: 2024
Type: article
Abstract: In radiotherapy, 2D orthogonally projected kV images are used for patient alignment when 3D-on-board imaging(OBI) unavailable. But tumor visibility is constrained due to the projection of patient’s anatomy onto a 2D plane, potentially leading to substantial setup errors. In treatment room with 3D-OBI such as cone beam CT(CBCT), the field of view(FOV) of CBCT is limited with unnecessarily high imaging dose, thus unfavorable for pediatric patients. A solution to this dilemma is to reconstruct 3D CT from kV images obtained at the treatment position. Here, we propose a dual-models framework built with hierarchical ViT blocks. Unlike a proof-of-concept approach, our framework considers kV images as the solo input and can synthesize accurate, full-size 3D CT in real time(within milliseconds). We demonstrate the feasibility of the proposed approach on 10 patients with head and neck(H\&N) cancer using image quality(MAE:<45HU), dosimetric accuracy(Gamma passing rate(2%/2mm/10%):>97%) and patient position uncertainty(shift error:<0.4mm). The proposed framework can generate accurate 3D CT faithfully mirroring real-time patient position, thus significantly improving patient setup accuracy, keeping imaging dose minimum, and maintaining treatment veracity.