Year: 2024
Type: article
Abstract: Aim: This study was aimed to calculate the working length (WL) of permanent teeth with simulated 45° resorption of the apical part of the root using four electric apex locaters (EALs): NSK, Woodpecker III, Woodpecker V, and Eighteeth. Methods: Twenty maxillary anterior single-rooted teeth were removed. Following tooth preparation to offer access to the root canal and replicate the 45° apical root resorption, each tooth underwent a microscope-assisted working length determination process. The four apex locaters measured each individual tooth and calculated electronic working lengths. Results: No discernible change existed between all four apex locater devices. Conclusion: NSK, Woodpecker III, Woodpecker V, and Eighteeth provide virtually the same measurement for 45 apical root resorption in single rooted teeth.
Source: Dentistry 3000
Institutions: University of Tikrit, Ministry of Health
Language: English
Subfield: Oral Surgery
Field: Dentistry
Domain: Health Sciences
Open Access status: diamond
OpenAlex ID: W4404624144