Clinical evaluation of the PowerGlide Pro midline catheter– dwell time, complications and outcomes for various medications including prostaglandins
Year: 2024
Type: article
Abstract: PURPOSE: The PowerGlide Pro™ Midline Catheter is a peripheral venous access device with a length of 8-10 cm, allowing the tip to reach far into the venous system. The aim of this study was to evaluate the dwell time of the catheter. Secondary endpoints included suitability for specific medications (e.g. prostaglandins) and assessment of complications. METHODS: Between January 2019 and November 2021, 50 patients were included in the study. Data on patient demographics, placement characteristics, complications and reasons for removal, were collected. RESULTS: Placement was technically successful in 92% (n = 46) of cases. In all cases, veins of the upper extremity were punctured (34 basilic veins, 7 brachial veins, 6 cephalic veins, and 3 median cubital veins). The average dwell time was 6.1 days (1-17 days). A significant difference between duration and medication administered could not be demonstrated. CONCLUSIONS: The longer maximum dwell time compared to a standard peripheral venous catheter makes it particularly suitable for intravenous therapy for more than 7 days or patients who have poor peripheral vein status.
Source: Langenbeck s Archives of Surgery
Authors: Yaser Souri, Edgar Franklin Hernandez Cancino, Hagen Kerndl, Alexander Hyhlik-Duerr, Yvonne Gosslau
Institution: University of Augsburg
Language: English
Subfield: Emergency Medical Services
Field: Health Professions
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $4,590
OpenAlex ID: W4404776797