Performance of the RadPro myOSL 4.0 whole body dosimeter
Year: 2024
Type: article
Abstract: Our whole-body personal dosimeter using the RadPro myOSL 4.0 system was purchased to replace an existing personal dosimeter system based on TLDs. For accreditation purposes it was tested against the Swiss requirements on personal dosimetry and some IEC 62387 requirements. Various irradiations compatible to the ISO-4037 definitions were carried out in the irradiation facility of the institute. An internally developed dose calculation was used to further improve the results. We found that our dosimetry system fulfills all Swiss and the tested IEC 62387 requirements for H p (10) and H p (0.07).
Source: Radiation Measurements
Author: A. Pitzschke
Language: English
Topic: Radiation Dose and Imaging
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $3,330
OpenAlex ID: W4404802205