Evaluation of Siemens Atellica AMH assay and comparison with Roche, Beckman and Ansh Labs
Year: 2024
Type: article
Abstract: • Siemens Atellica AMH assay showed minimal bias and strong correlation with DxI. • Siemens Atellica assay exhibited biases versus Roche and Ansh Labs at high AMH levels. • A commutable standard and assay calibration transparency are vital for AMH testing. Anti-Müllerian hormone (AMH) is an important marker for ovarian reserve and response to fertility treatments. However, interassay variability exists due to the lack of a standardized method. This study evaluates the performance of the new Siemens Healthineers Atellica IM AMH assay against established assays (Beckman DxI 600 Access, Roche cobas Elecsys® e801, and Ansh Labs AMH ELISA). We measured AMH concentrations in 120 residual serum samples from patients presenting for routine AMH testing using all four assays. Passing-Bablok regression and Bland-Altman method were used for data analysis. The Siemens Healthineers Atellica IM assay demonstrated strong correlation with Beckman DxI (slope: 1.07, R 2 : 0.9881) but showed a minimal positive bias. Conversely, the Roche cobas Elecsys® e801 assay exhibited a negative bias compared to Beckman DxI (slope: 0.74, R 2 : 0.9696), while the Ansh Labs assay demonstrated a significant positive bias with increasing variability at higher AMH concentrations. Overall, the Siemens Healthineers Atellica IM assay shows promise as an alternative method for AMH measurement, demonstrating good correlation with established assays. However, differences were observed between all assays, highlighting the importance of assay-specific interpretation for accurate clinical assessment.
Source: Clinical Biochemistry
Authors: Emily Shang, Hilario Agustin, Jason K.Y. Lee, Matt Scheidegger, JoAnna Jones, Charbel Abou-Diwan, Joe M. El-Khoury, Jieli Li
Language: English
Subfield: Physiology
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $3,000
OpenAlex ID: W4405364644