Work

Evaluating Piezoelectric Ballistocardiography for Post-Surgical Heart Rate Monitoring

Year: 2024
Type: conference-paper
Abstract: Ballistocardiography (BCG) is a non-invasive technique measuring heart and circulatory activity, proposed as an alternative to ECG for long-term heart rate monitoring.This study evaluates a custom multi-node piezoelectric BCG sensing mat for heart rate measurement compared to single-lead ECG in post-cardiac surgery patients.14 patients (10 males, mean age 63.47 ± 9.35) postcoronary artery bypass (N=8) or heart valve repair (N=6) were enrolled at Maastricht University Medical Center.Patients in the cardiac ICU used a bed with a custom BCG sensor mat (10 PVDF-TrFE piezoelectric sensors, 90x50 cm²).BCG and ECG were recorded simultaneously for an average of 15 hours per patient.BCG heart rate values were extracted over ten-second epochs using a multinode rule-based algorithm.Performance was assessed using mean absolute error (MAE) comparing BCG and ECG heart rates.Heart rate values derived from the piezoelectric BCG mat resulted in a weighted mean MAE of 1.41 bpm with 66.14% coverage compared to ECG.Piezoelectric BCG shows promise as an accurate heart rate measurement method in post-cardiac surgery patients, offering a non-invasive alternative to ECG.These findings support further development of this technology for clinical use in remote heart rate monitoring.
Language: English
Open Access status: bronze
OpenAlex ID: W4405434827
FWCI:
Cites:
Cited by:
Related to: 10