Patterns of pulsatile impedance change during the apnea test detected by electrical impedance tomography

Year: 2024
Type: article
Abstract: The apnea test (AT) is a critical procedure in determining brain death. It temporarily disconnects the patient from the ventilator for 7 to 10 minutes. This study explores pulsatile Electrical Impedance Tomography (EIT) during AT in 10 patients suspected of brain death to discern physiological changes. Analysis revealed three distinct patterns of pulsatile EIT signal changes: increasing; fluctuating; and decreasing. These findings show the potential of pulsatile EIT monitoring to provide novel insight into cardio-pulmonary dynamics during AT.