Abstract: A new method was proposed for calibrating large electromagnetic flow meters using small flowrates. The method has traceability and can be theoretically supported. It can avoid using large volume flow rigs and much power consumption. Preliminary experiment was done. Theoretical analysis was given briefly. Uncertainty of the method was discussed.