Efficient 3D FISP-MRF at 0.55 T using long spiral readouts and concomitant field effect mitigation
Year: 2025
Type: article
Abstract: PURPOSE: To demonstrate the feasibility of SNR-efficient 3D fast imaging with steady state precession MR fingerprinting (FISP-MRF) using long spiral readouts with mitigation of concomitant field effects at 0.55 T. METHODS: Fourteen FISP-MRF sequences with different spiral readout lengths (2.9 ms to 22.0 ms) were implemented with the open-source Pulseq framework. Datasets were reconstructed using a low-rank and subspace model-based reconstruction combined with MaxGIRF spatial encoding model. Concomitant field-induced blurring and MRF precision were evaluated using reconstructed images and quantitative maps acquired from an ACR phantom, an ISMRM/NIST system phantom, and 2 healthy volunteers. RESULTS: standard deviations of in-vivo white matter were reduced by ∼50 %, and concomitant field mitigation was necessary. Residual spatial blurring was non-negligible for readouts ≥16.5 ms, suggesting an operating regime (2.9-16.5 ms) for 3D FISP-MRF at 0.55 T. CONCLUSION: We demonstrate SNR-efficient 3D FISP-MRF using long spiral readouts in conjunction with concomitant field-induced blurring mitigation. A wide operating regime with improved precision is feasible only after concomitant field mitigation.
Source: Magnetic Resonance Imaging
Authors: Zhibo Zhu, Nam G. Lee, Krishna S. Nayak
Institution: University of Southern California
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $3,380
OpenAlex ID: W4407607429
