A simplified γ-γ coincidence method to measure the activity of radionuclide
Year: 2025
Type: article
Abstract: A simplified γ-γ coincidence method for activity measurement when a sample containing multiple γ-ray emitters is proposed. This method can be applied when a γ-γ cascade process occurs in the radionuclide of low detection efficiency. And the activity can be calculated from the full-energy peak count rate of the cascade particles in each detector and the corresponding full-energy-full-energy coincidence count rate. To verify this simplified method, it was compared with numerous γ-γ coincidence methods using Monte Carlo simulation . Limitations and applicability of these methods are discussed. In addition, to validate this method in real environments, a digital coincidence measurement system using two NaI(Tl) detectors was developed. The measurements were performed on a standardized 60 Co source and a standardized 60 Co source in the presence of a 137 Cs standardized source.
Source: Radiation Measurements
Authors: Qiankun Shao, Qingjun Zhu, Jia Li, Songlin Liu
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: hybrid
Estimated APC paid: $3,690
OpenAlex ID: W4407834387