Abstract: A multi-gas sensing system based on mid-infrared spectral absorption was developed for the detection of CH4, C2H6, and C2H4. The system utilized a broadband infrared source, a hollow waveguide (HWG) absorption cell, and a tunable Fabry–Pérot (FP) detector. The limits of detection (LODs) of CH4, C2H6, and C2H4 were 7.33 ppm, 2.13 ppm, and 8.09 ppm, respectively. For multi-gas measurements, the support vector machine (SVM) algorithm model was employed to calculate the concentration of each component. The root mean square error of prediction (RMSEP) values for CH4, C2H6, and C2H4 were 15.91 ppm (1.26%), 1.64 ppm (0.57%), and 6.95 ppm (0.55%), respectively. The generation of stimulated absorption spectra of mixed gases was realized, and the sample selection of measurement for accurate concentration calculation of each gas was optimized. The system proposed in this work provides a simple, miniaturized, and cost-effective solution for multi-gas sensing.
Source: Sensors
DOI: 10.3390/s25051427
Authors: Wenyuan Shao, Yunjiang Jia, Xilian Su, Benlei Zhao, Jiachen Jiang, Limei Gao, Xiaosong Zhu, Yiwei Shi
Institution: Fudan University
Language: English
Subfield: Spectroscopy
Field: Chemistry
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $3,137
Award: G3769601387
OpenAlex ID: W4407955411