Work

Device for dynamic calibration of pressure sensors

Year: 2024
Type: article
Abstract: Modern devices for dynamic calibration of pressure sensors (shock tubes, power simulators of pressure impulse, etc.) have a number of drawbacks stemming from the principles of creating a test impact. Besides, the problem of rational choice of the method of calibrating pressure sensors depending on the dynamic parameters of the sensor and the required test accuracy has not been solved for modern test systems. The paper presents a solution to the problem of correlation between the test parameters, dynamic parameters of the pressure sensor and test accuracy. The obtained analytical dependencies of such a relationship make it possible to reasonably select or develop a method for studying the dynamic characteristics of sensors. Based on theoretical studies, the principle of creating a test impact has been proposed, and a method and device for implementing dynamic calibration of pressure sensors have been devised. The developed device allows the transient response of the sensor to be obtained, as well as setting the decay time of its natural vibration. Based on the transient response, other dynamic characteristics of the sensor, namely the impulse transient and frequency response, can be calculated.
Language: English
Open Access status: diamond
OpenAlex ID: W4407991464
FWCI: 0.515
Cites:
Cited by: 3
Related to: 10