Abstract: The beam modeling of the linear accelerator is performed based on dosimetric measurements by including the data obtained for the absorbed dose in water at a depth of 10cm (D10), percent depth doses, and geometric symmetry of the fields. The results obtained for photon beams with the energy 6MV and 18MV are placed in a database of the treatment planning system (XiO). Then, it becomes a reference system for modeling the photonic beams with high energy and creating patient treatment plans. The field modeling and calculation of the dose that should be given to the patient are done by precise dosimetric measurements.
Authors: Ervis Telhaj, Blerim Rrakaqi, Besim Xhafa, Njomza Elezaj, Ylli Kaçiu
Institutions: University Clinical Center of Kosovo, University of Prishtina
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green