Microfluidic Device on Fused Silica for Raman Spectroscopy of Liquid Samples
Year: 2025
Type: article
Abstract: Water testing is becoming increasingly important due to dangerous phenomena such as Harmful Algal Blooms (HABs). Commonly, the content of a water sample is measured for the detection, monitoring and control of these events. Raman spectroscopy is a technique for the molecular characterization of materials in solid, liquid or gaseous form, which makes it an attractive method for analysing materials' components. However, Raman scattering is a weak optical process and requires an accurate system for detection. In our work, we present, from design to fabrication, a microfluidic device on fused silica adapted to optimise the Raman spectrum of liquid samples when using a Raman probe. The device features a portable design for rapid on-site continuous flow measurements avoiding the use of large, costly and complex laboratory equipment. The main manufacturing technique used was ultrafast laser-assisted etching (ULAE). Finally, the effectiveness of the microfluidic device was demonstrated by comparing the Raman spectra of a known species of cyanobacteria with those obtained using other conventional substrates in laboratory analysis. The results demonstrate that the microfluidic device, under continuous flow conditions, exhibited a lower standard deviation of the Raman signal, reduced background noise and avoided signal variations caused by sample drying in static measurements.
Source: Biosensors
DOI: 10.3390/bios15030172
Language: English
Subfield: Biophysics
Domain: Life Sciences
Open Access status: gold
Estimated APC paid: $2,654
OpenAlex ID: W4408186828