First prospective, single-arm, multicenter study to evaluate safety and efficacy of the overall thrombectomy system -iNedit, iNdeep, and iNtercept- for acute ischemic stroke. Rationale beyond the study

Year: 2025
Type: article
Abstract: Rationale: The clinical impact of a novel mechanical thrombectomy strategy, which integrates distal access with flow reversal and flow arrest via a distal balloon, all within a single procedure [Safety and Efficacy of the overall throMbectomy system for sTroke (SEMTiC) strategy], has not been tested. Aim: study evaluating the safety and efficacy of the combined thrombectomy system-iNedit, iNdeep, and iNtercept-in patients with acute ischemic stroke. Sample size estimates: The study was designed with a sequential structure based on the efficacy endpoint (eTICI ≥2b) reported in the literature [71.1% with a 95% confidence interval of (68.5%, 73.8%)]. An interim analysis was set for 115 patients and a final analysis for 225 patients, ensuring 98% power at a one-sided 0.025 significance level, with a 2.6% non-inferiority margin and a 15% assumed withdrawal rate. Design: SEMTiC-01 is a prospective, multicenter, single-arm, open-label clinical safety and efficacy investigation. Outcome: Primary efficacy endpoint: expanded treatment in cerebral infarction score (eTICI) ≥2b revascularization within ≤ 3 stent retriever passes. Primary safety endpoint: monitoring serious adverse events within 24 h post-intervention and all-cause mortality at 90 days.
Language: English
Subfield: Epidemiology
Field: Medicine
Open Access status: gold
Estimated APC paid: $3,801
OpenAlex ID: W4408219494
FWCI: 0.623
Cites: 20
Cited by: 1
Related to: 10