CFD and PIV analysis of a novel, low shear stress intra-aortic ventricular assist device
Year: 2025
Type: article
Abstract: Stroke has emerged as the primary contributor to morbidity and mortality in patients undergoing treatment with Left Ventricular Assist Devices (LVADs), possibly arising from the turbulent flow and elevated wall shear stresses generated in these devices. A minimally invasive LVAD (LifeheART) has been proposed to address these issues, employing an intra-aortic location and a shaftless impeller design. The current study uses Particle Image Velocimetry (PIV) flow visualization, carried out in a Cardiovascular Mock Circulation Loop (CMCL), to identify the velocity distribution at the pump outlet in order to validate the developed CFD model. Subsequently, the model evaluates the blood shear stress distribution and blood damage index. The results showed that the calculated viscous shear stress (VSS) and the blood damage index of the LifeheART prototype is significantly lower than the published data for current clinically available devices, confirming the potential utility of the new design to improve patient outcomes.
Source: Journal of Biomechanics
Authors: Elif Oran, Essam Abo-Serie, James Jewkes, Manus Henry, Bulent Oran
Institutions: Coventry University, University of Leicester, University of Oxford
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: hybrid
Estimated APC paid: $3,650
Award: G2989991374
OpenAlex ID: W4408327759