A 0.05 mm3 diode-based single charged-particle real-time radiation detector for electron radiotherapy
Year: 2025
Type: article
Abstract: Real-time radiation monitoring at the single-particle level is an unmet need for electron radiotherapy, especially for dose deposition to targets in motion or critical OARs. We have developed a first-in-class CMOS-based 0.05 mm 3 single electron sensitive detector. The chiplet integrates all the requisite electronics. The functionality of the system is verified under 6 and 9 MeV clinical electron beams. Percentage depth vs. pulse-width curves for 6 and 9 MeV beams are measured and verified using Monte-Carlo simulations. The proposed system has the potential to enhance the electron radiotherapy quality and safety, providing real-time dosimetry from multiple sites simultaneously.
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $1,980
Awards: G3649934610, G4518611287, G5407469067