Spot-optimization reduces beam delivery time in liver breath hold intensity modulated proton therapy
Year: 2025
Type: article
Abstract: Background and purpose: Liver irradiations with intensity-modulated proton therapy (IMPT) often require motion mitigation techniques that prolong treatment. A prototype spot-optimization algorithm was tested to evaluate whether plan delivery time could be reduced while preserving quality. Methods and materials: Fifteen patients previously treated with liver IMPT using breath-hold were re-planned with nominal treatment planning system (TPS) settings and using a prototype spot-optimization algorithm in which combinations of minimum Monitor Unit (MU) and layer-spacing settings were tested: 1MU/1MeV, 3MU/3MeV, 1MU/5MeV, 5MU/3MeV. Spot-optimized and nominals plans were compared using standard dose-volume histogram (DVH) metrics for targets and organs-at-risk. A Wilcoxon signed-rank test was applied (p < 0.05). Delivery time for all plans were measured by creating and delivering IMPT quality assurance (QA) plans. Gamma analyses were performed on all plans to test deliverability. Plans were considered deliverable if >90 % of points passed a gamma criterion of 3 %/3mm. Results: 0.4 Gy(RBE)). QA analysis showed all spot-optimized plans were deliverable. Conclusions: The spot-optimization algorithm produced clinically deliverable plans with negligible DVH differences to nominal plans and reduced delivery time of liver IMPT by over one-third.