On-site salt iodine detector (OSID) kit: An alternative tool for salt iodization monitoring
Year: 2025
Type: article
Abstract: = 0.9956) with limit of detection (LOD, 3.6 mg/Kg), method recovery (89.5 %-99.4 %), intra-assay imprecision (coefficient of variation, CV 5.7 %), inter-assay imprecision (CV 7.0 %) and inter-operator imprecision (CV 6.1 %). A 30.0 mg/Kg iodate and sample S4 used as quality control (QC) for ten days monitoring which resulted within the range of mean±2×standard deviation (SD) (27.2 ± 2.4 mg/Kg and 21.6 ± 5.3 mg/Kg, respectively). The OSID kit showed strong agreement in a Blant-Altman plot when compared with UHPLC (60 salt samples) indicating a reliable and cost-effective alternative for salt iodine monitoring.•The optimization of salt weight and volume of solution D allows 100 analyses per kit.•The OSID kit imprecision validation of <10 % demonstrate excellent precision and reliability of the method.•The result comparison between the OSID kit and the UHPLC reference method suggest a strong agreement in a Blant-Altman plot.
Source: MethodsX
Authors: Mohd Azerulazree Jamilan, Nurul Aznyda Norizan, Ahmad Ali Zainuddin, Noor ul-Aziha Muhammad, Filza Noor Asari, Aswir Abd Rashed, Mohd Fairulnizal Md Noh
Language: English
Subfield: Endocrinology, Diabetes and Metabolism
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $1,260
Funder: Kementerian Kesihatan Malaysia
Awards: G3667838710, G389623200
OpenAlex ID: W4409413911
